Abstract
The authors describe a method for performance of ultrafast magnetic resonance (MR) angiography of coronary arteries with a standard clinical MR system and a body coil. Each image was obtained within a single breath hold by using an electrocardiography-gated, segmented, ultrafast, gradient-echo pulse sequence with an incremental excitation flip angle for the eight phase-encoding steps acquired per segment. By using overlapping 4-mm-thick sections, the coronary arteries were routinely depicted from the coronary ostia distally at MR in healthy subjects. Ultrafast MR angiography of the coronary arteries is feasible with use of a standard body coil. This technique offers considerable potential as an investigational tool and, with further development, may become a clinically useful imaging application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.