Abstract

A corona discharge (CD) ionization source was prepared for a planar high-field asymmetric waveform ion mobility spectrometer (FAIMS). The effects of discharge current and discharge distance on ionization efficiency were investigated; and the electric field dependence of the ion injection in the reaction region was studied. The results showed that the discharge current of CD source had good linearity with the intensity of reactant ion peak (RIP), and the RIP intensity increased to a stable level at the discharge distance of >5 mm. An injection electrode was introduced to improve the ionization efficiency. A square-wave voltage applied to the electrode was found to provide optimal performance of ion injection and utilization. The operating parameters of the CD-FAIMS were optimized to achieve trace level detection of dimethyl methylphosphonate (DMMP) sample. The detection limit for DMMP was 0.5 µg/m3.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.