Abstract
The central corneal zone is depicted on keratoscope photographs using a small target aperture and a large object distance. Information on the peripheral area is included by employing a hemispherical target with a dense circular and radial pattern. On a 16 mm (R = 8 mm) reference steel sphere the diameter of the innermost reflected mire is 0.20 mm and the distance between mires 0.13 mm. The standard deviation on focusing precision, determined to 0.20 mm employing a fixed object, corresponds to an error of 0.06% in linear measurements on keratographs. The variation in object distance due to the curved image plane is at most 0.55 mm within 60 deg from the optical axis of the system for the reference sphere. Nevertheless, all reflected mires are distinctly reproduced on the keratoscope photograph. For a test object the slopes of light intensity curves on the passage between black and white areas are reduced by only 1.0% with a variation in object distance of 0.55 mm around focus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.