Abstract

Previously, we reported that the Ex-press® shunt (EXP) was associated with more rapid reduction in corneal endothelial cells when inserted into the cornea rather than the trabecular meshwork (TM). We compared the reduction rate of corneal endothelial cells between the corneal insertion group and TM insertion group. This was a retrospective study. We included patients who had undergone EXP surgery and were followed for > 5years. We analyzed the corneal endothelial cell density (ECD) before and after EXP implantation. We included 25 patients in the corneal insertion group and 53 patients in the TM insertion group. One patient in the corneal insertion group developed bullous keratopathy. The ECD decreased significantly more rapidly in the corneal insertion group (p < 0.0001), in whom the mean ECD decreased from 2227 ± 443 to 1415 ± 573cells/mm2 at 5years with a mean 5-year survival rate of 64.9 ± 21.9%. By contrast, in the TM insertion group, the mean ECD decreased from 2356 ± 364 to 2124 ± 579cells/mm2 at 5years, and the mean 5-year survival rate was 89.3 ± 18.0%. The decrease rate of ECD was calculated as 8.3%/year in the corneal insertion group and 2.2%/year in the TM insertion group. Insertion into cornea is a risk factor for rapid ECD loss. The EXP should be inserted into the TM to preserve the corneal endothelial cells.