Abstract
We report on a novel concept and prototype development of a coreless SQUID-based charged-particle beam monitor as a non-destructive diagnostic tool for accelerator facilities. Omitting the typically used pickup coil with a high magnetic permeability core leads to a significant improvement in low-frequency noise performance. Moreover, a revised shielding geometry allows for very compact and rather lightweight device designs. Based on highly sensitive SQUIDs featuring sub-micron cross-type Josephson tunnel junctions, our prototype device exhibits a current sensitivity of about 6 pA Hz−1/2 in the white noise region. Together with a measured shielding factor of about 135 dB this opens up the way for its widespread use in modern accelerator facilities.