Abstract

A new method for the analysis of triglycerides (TAGs) in vegetable oils was developed using a Poroshell 120 EC-C18 column (3.0mm×50mm, 2.7μm) with a high resolution ESI-Q-ToF tandem mass spectrometer as detection system. We used an Agilent Poroshell column, which is characterized by a recently developed stationary phase based on non-porous core particles. The results highlighted the advantages of this column in terms of the dramatic improvement in the number of theoretical plates and in low column backpressure. The developed method enabled us to analyze complex mixtures of more than 40 TAGs within less than 25min and with a low backpressure (lower than 100bar), and represents the first application of a core-shell stationary phase in reverse phase HPLC using an ESI-Q-ToF as detection system. The method was optimized on standards of TAGs, validated and applied to several plant oils. By a quantitative point of view, the method showed a very good linearity (r2>0.999) in the range 0.1–2.4μg/g; high intra- and inter-day precision both in terms of retention times (RSD%<0.04%) and peak areas (RSD%<0.3%). Limits of detection and quantitation were lower than 0.03μg/g and 0.10μg/g, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.