Abstract

A TiO(OH)2 precipitate was used for the preconcentration of copper(II), zinc(II), and lead(II) in seawater prior to determination by flame atomic absorption spectrometry. The influence of pH, sample volume, amount of precipitate, and centrifugation time were optimized for quantitative recoveries of the analytes. Under the optimum conditions, the detection limits of copper(II), lead(II), and zinc(II) were 4.3, 9.7, and 9.6 micrograms per liter, respectively. The recoveries of analytes were between 95.00 and 103.00 percent with the relative standard deviation below 6 percent. The procedure was validated by the analysis of NASS-5 and SPS-WW1 Batch 109 standard reference materials and the procedure was successfully applied to seawater.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.