Abstract
Polycrystalline mercuric iodide films are being developed as a new detector technology for X-ray imaging at room temperature. In this paper, different electrode patterns were designed by varying the widths of grid and gap. The finite elements analysis was used to simulate the potential distribution of the coplanar-grid detectors. The electrode design of pattern A (the width of grid was 150μm and the width of gap was 500μm) was the best. In addition, the influence of edge effects resulting from the asymmetrical electrodes on the potential distribution was discussed. In this way, the geometry of electrodes for coplanar-grid detectors was optimized by widening the second grids near the edges as shown in pattern N.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.