Abstract

Conversion factors of exposure to absorbed dose f in compact bone, muscle, water and adipose tissue are deduced from the measured photon fluence spectra of X-ray beams with potentials from 50 to 250 kV. The results show that accurate determination of f requires beam quality specification not only by the HVL value but also by a second parameter, e.g. the peak tube potential. The differences between the conversion factors f obtained in this work and those deduced from the monoenergetic equivalent energy are evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.