Abstract

The purpose of this experimental study was to develop an alternative technique of arterial microanastomosis using only 2 stay-sutures augmented with fibrin glue, and to compare it to the conventional technique in arteries of varying diameters mimicking hand arteries. Eight anastomoses were performed in 7 male rats, including 1 anastomosis each on the 2 femoral, iliac, and carotid arteries, and 2 on the subrenal aorta. The conventional technique was used on one side and on the first aorta anastomosis, while augmented anastomoses were performed on the other side and on the second aorta. Patency was tested 10 min after unclamping; clamping time, blood loss, anastomosis quality score (out of 15 points) and artery diameter were recorded. In arteries of diameter 0.5−2.2 mm, augmented anastomoses were on average 10.7 ± 3.2 min faster to perform (p < 0.0001), with an average of 1.3 ± 0.9 g less blood loss (p < 0.0001) and an average of 2.6 ± 2.5 points higher quality score (p < 0.0001). There were no significant differences between the two techniques in terms of patency rate, regardless of artery size. However, 3 of the 7 augmented anastomoses were non-permeable in the femoral subgroup (i.e., submillimetric arteries). This straightforward technique appears to be time-saving and reliable, provided that the repaired artery is of sufficient size. Subject to clinical validation, this technique might help surgeons treating extensive hand wounds with multiple severed neurovascular bundles.