Abstract

The quantification of venous flows can be obtained by multiplying cross sectional area, measured on a B-mode video-clip, by velocity values, obtained with Doppler measurements. The Doppler angle between ultrasound (US) line and blood flow requires a manual adjustment. Establishing this parameter is critical in order to calculate blood velocity. However, the operator dependency gives high variable results. It is worth noting that a new class of vector Doppler devices can enhance the accuracy and precision of measurements. Such technology uses a double US line that leads to automatically know the Doppler angle. By comparing in an <em>in vitro</em> model of venous flow conventional echo color Doppler (ECD) equipment with the new device, we found a better minimal difference between the latter and the nominal flow rate (20%). On the contrary, the comparison with conventional ECD showed a difference ranging between 2% and 43%, according to the possible settings of the equipment. Our study demonstrates a better accuracy of the experimental device with respect to conventional ECD in measuring the venous flow rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.