Abstract

We developed a conventional immunosensor for fibrinogen and fibrin degradation products (FDP) to combine a quartz crystal microbalance (QCM) with the agglutination reaction of immunized latex beads. FDP induced an immunoreaction due to anti-FDP antibody immobilized latex particles. We successfully measured FDP concentration of in human serum within 10 min by QCM method. The detection range of QCM immunosensor is covered with screening concentration of FDP in serum (<10 μg/ml of FDP). The time course of latex agglutination obtained from QCM immunosensor is synchronized to that of latex photometric immunoassay. SEM was used to observe the surface of QCM that applied FDP serum. The size of latex particles agglutinated on the QCM electrode increased concomitant with FDP concentration. Frequency shift on immunoreaction explains the increased adsorption amount of agglutinated latex on QCM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call