Abstract

A method for the determination of trace elements in solid samples is described. The sample is loaded into a graphite cup, which is inserted horizontally into the tail region of the inductively coupled plasma. Application of the method to carrier distillation analysis of impurities in uranium oxide shows that detection limits in the range 0.02–5 µg g–1 with a relative standard deviation of 4–12% are obtainable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.