Abstract

We applied various prepurification protocols (extraction with different solvents, liquid/solid separation on bonded silica media, Celite, and Sephadex LH20 chromatography) with a range of commercially available RIA kits to measure cortisol in urine samples. We then compared the results with the concentrations measured by a HPLC method validated with reference to isotope dilution gas chromatography-mass spectrometry. We conclude that chromatography on a commercial, prepacked diol minicolumn (Waters Sep-Pak Vac RC) in combination with dichloromethane extraction is a convenient and very effective purification step before RIA of urinary cortisol in patients not receiving corticoid medication. We tested numerous steroids for interference and found that free polar cortisol derivatives (hydroxylated or hydrogenated) could only partially account for the overestimations routinely encountered when free urinary cortisol concentrations are measured by direct RIA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.