Abstract
On-line voltammetric measurements under liquid chromatographic (LC) conditions yield voltammograms which are either purely diffusion- or purely convection/diffusion-controlled, or a combination of both. The dependence of this behaviour on flow-rate, scan-rate and electrode diameter is investigated for a large-volume wall-jet detector. It is shown that for 4 V s −1 scan-rates at 1 ml min −1 flow-rates (conventional LC conditions), S-shaped (convection/diffusion-controlled) voltammograms can be obtained with macro-electrodes (⩾ 1 mm diameter). Distortion of voltammogram shape by the cell time constant is discussed for macro-electrodes. The behaviour of the cell in microbore and micro-LC applications is demonstrated. The advantage of being able to change from convection/diffusion-controlled to diffusion-controlled behaviour is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.