Abstract

A theoretical analysis has been performed to show the interrelationships between alveolar carbon dioxide concentration, carbon dioxide output, alveolar ventilation and fresh gas flow during controlled ventilation with the Mapleson D system. The model and forms of equation used have been substantiated by experiments in which dogs were ventilated through a coaxial version of the Mapleson D system at varying levels of fresh gas flow and alveolar ventilation. By assuming that man ventilated with this system behaves as does the dog, a nomogram has been produced to predict alveolar carbon dioxide concentrations at any levels of fresh gas flow and minute volume.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.