Abstract

Total plutonium is selectively determined by a controlled-potential coulometric method in which plutonium is reduced to Pu(III) at 0.25 V (vs. SCE) in a 5.5 M hydrochloric acid—0.015 M sulfamic acid electrolyte, diverse ions are oxidized at 0.57 V, phosphate is added to reduce the Pu(III)—Pu(IV) potential, and Pu(III) is oxidized to Pu(IV) at 0.68 V. None of more than 50 metal ions present in nuclear fuel-cycle material interferes. Many anions are without effect and most interfering ones are removed by preliminary fuming with perchloric acid. The apparatus described consists mainly of commercial components. The relative standard deviation is <0.1% at the 5-mg plutonium level.