Abstract

We aimed to develop a control system for multichannel magnetocardiography (MCG) based on a high-temperature DC superconducting quantum interference device (high-Tc SQUID). To create this system, we used one oscillator as an AC bias controller to operate a multichannel high-Tc SQUID. To optimize the SQUID parameters (such as the AC bias, offset voltage), two new control sequences based on a cross-correlation method and a fast Fourier transform method were developed. Using the AC bias controller and the sequences, the typical white noise level of the SQUID was about 50–60 fT Hz-1/2 around 100 Hz. Multichannel MCG signals were detected clearly in the system with the SQUIDs. We conclude that our control system with one oscillator and new protocols can reliably operate a multichannel SQUID.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call