Abstract

At present, such a type of endoscopic surgery as robot-assisted transaxillary thyroidectomy has become available. In this case, traumatization of the brachial plexus is a rare but possible complication. For the control of the function of the brachial plexus during the above operation we used monitoring of somatosensory evoked potentials (SSEPs) induced by stimulation of the median nerve. Fifteen patients (14 women and one man) were included in this study. All interventions were robot-assisted transaxillary thyroidectomies using the daVinci SI Surgical System. We found that such surgery induced mild but significant increases in the latency of the cortical N20 potential, especially when the tissue was tensioned by the autostatic retractor. The latency prolongation was a valuable signal given to the surgeon, followed by repositioning or loosening of the retractor. In the examined group, no significant decreases in the amplitude of the N20 potential were observed. We conclude that SSEP monitoring during robotic thyroid surgery is an available and safe method providing valuable information on the functional integrity of somatosensory pathways during surgical maneuvers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.