Abstract
The problem of embedding sensor fault tolerance in feedback control of neuromuscular blockade is considered. For tackling interruptions of feedback measurements, a structure based upon Bayesian inference as well as a predictive filter is proposed. This algorithm is general and can be applied to different situations. Here, it is incorporated in an adaptive automatic system for feedback control of neuromuscular blockade using continuous infusion of muscle relaxants. A significant contribution consists in the experimental clinical testing of the algorithm in patients undergoing surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.