Abstract
BackgroundIatrogenic inferior vena cava (IVC) injury is a rare but potentially life-threatening complication during laparoscopic surgery. This experimental study aimed to assess the hemostatic ability of a new device, double balloon-equipped central venous (DB-CV) catheter, for IVC injury.MethodsThe DB-CV catheter comprises a triple-lumen sphincterotome combined with two dilating balloons having a diameter of 25 mm. The experimental procedures were performed in five pigs. The DB-CV catheter was inserted via the right femoral vein. For the IVC occlusion test, correct placement of the balloons was confirmed by indocyanine green fluorescence imaging, and hemodynamic data were recorded. For the IVC injury test, a 3- to 4-mm circumferential incision was created in IVC, and hemostasis was initiated using balloon inflation 5 s after the injury.ResultsHemodynamic changes were minimal, with a 20 mmHg reduction in the mean arterial pressure because of IVC occlusion. All bleeding from IVC injuries was successfully temporarily stopped by direct balloon compression, with a mean time to hemostasis of 69 s and mean blood loss of 32 ml. Subsequently, the positioning of IVC injuries between two balloons made it possible to suture the injured IVC.ConclusionsBalloon occlusion using the DB-CV catheter provides a rapid temporal hemostatic effect and can overcome the serious condition of massive hemorrhage from IVC injuries.
Highlights
Iatrogenic inferior vena cava (IVC) injury is a rare but potentially life-threatening complication during laparoscopic surgery
It is highly recommended to convert to an open surgery if IVC injury occurs
During emergency laparotomy, several minutes are required to reach the injury site, and further blood loss may result from removal of pneumoperitoneum pressure
Summary
Iatrogenic inferior vena cava (IVC) injury is a rare but potentially life-threatening complication during laparoscopic surgery. This experimental study aimed to assess the hemostatic ability of a new device, double balloonequipped central venous (DB-CV) catheter, for IVC injury. For the IVC occlusion test, correct placement of the balloons was confirmed by indocyanine green fluorescence imaging, and hemodynamic data were recorded. For the IVC injury test, a 3- to 4-mm circumferential incision was created in IVC, and hemostasis was initiated using balloon inflation 5 s after the injury. All bleeding from IVC injuries was successfully temporarily stopped by direct balloon compression, with a mean time to hemostasis of 69 s and mean blood loss of 32 ml. The positioning of IVC injuries between two balloons made it possible to suture the injured IVC
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have