Abstract

Purpose Measurement of absorbed dose in target volume is widely considered to be an important tool for quality assurance in external radiotherapy. The aims of this work were to measure the entrance dose for patient treated for head and neck tumors and to compare this measured dose with the dose calculated. Patients and methods Twenty patients were evaluated. Initially, the measurements were performed on a polystyrene phantom in order to calibrate diodes in terms of entrance dose and to determinate appropriate correction factors. In vivo entrance dosimetry check was performed for these patients treated for head and neck tumors in 60Co gamma-rays. Results For the entrance dose evaluation over 100 field measurements, the mean deviation between the measured dose and the calculated dose was equal to 0.12% and the standard deviation was 1.84%. The deviation was less than 3% in 95% of measurements. Large deviation (more than 5%) was observed in one case. Conclusion Simple in vivo dose measurements are an additional safeguard against major set-up errors and calculation or transcription errors that were missed during pretreatment chart check.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.