Abstract We report on investigations of how fast changes of the refractive index influence the uncertainty of interferometric displacement measurements. Measurement of position within a limited range is typical for precise positioning of coordinate measuring systems, such as nanometrology standards combined with scanning probe microscopy (SPM). The varying refractive index of air contributes significantly to the overall uncertainty; it plays a role especially in case of longer-range systems. In our experiments we have observed that its fast variations, seen as length noise, are not linearly proportional to the measuring beam path and play a significant role only over distances longer than 50 mm. Thus, we found that over longer distances the length noise rises proportionally. The measurements were performed under conditions typical for metrology SPM systems
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.