Abstract

The authors aimed to elucidate the factors related to effective lens position, tilt, and decentration of scleral fixed intraocular lenses (IOLs) with a flanged haptic technique in an artificial eye model using anterior segment optical coherence tomography. Two bent 27-gauge needles were passed through a 1.0- or 2.0-mm scleral tunnel, 2.0 mm posterior to the limbus and 180° apart. Both haptics of a three-piece IOL were docked with guide needles and externalized. Factors related to the IOL position were analyzed using anterior segment optical coherence tomography and a stereomicroscope. The 1.0-mm scleral tunnel induced a significantly longer effective lens position than the 2.0-mm tunnel and suture fixation ( P < 0.05 and P < 0.01, respectively). Discrepancy in scleral tunnel length induced higher decentration of the optic to the opposite side of the haptic-embedded shorter tunnel and tilt perpendicular to the fixed axis than that in the scleral tunnel of the same length ( P < 0.001 and P < 0.05, respectively). If the scleral fixation points of both haptics are not exactly 180° apart, the IOL may become decentered and tilted ( P < 0.01 and P < 0.05, respectively). In the flanged haptic technique, the length, balance, and position of both scleral tunnels determine IOL effective lens position, tilt, and decentration.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call