Abstract

In a perfect optical system numerical aperture and wavelength determine resolution. In a real optical system, however, the number of photons collected from a specimen determines the contrast and this limits the resolution. Contrast is affected by the number of picture elements per unit area, the number of photons and the aberrations present in every optical system. The concept of contrast vs. distance functions is used to compare the resolution achievable in confocal and wide‐field fluorescence microscopes and the effect of a further reduction of the observable volume. In conclusio: (a) real optical systems will never be able to achieve the theoretical resolution, (b) wide‐field fluorescence microscopy will often provide a better resolution than confocal fluorescence microscopy, (c) decreasing the observed volume does not necessarily increase the resolution and (d) using multiple fluorophores can improve the accuracy with which distances are measured. Some numbers for typical situations are provided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.