Abstract

BackgroundAccurate evaluation of the tricuspid regurgitant (TR) spectral Doppler signal is important during transthoracic echocardiographic (TTE) evaluation for pulmonary hypertension (PHT). Contrast enhancement improves Doppler backscatter. However, its incremental benefit with contemporary scanners is less well established. The aim of this study was to assess whether the TR spectral Doppler signal using contemporary scanners was improved using a second generation contrast agent, Definity® (CE), compared to unenhanced TTE (UE). MethodsAnalysis of patients who underwent UE then CE TR interrogation was performed. TR signal was evaluated by an experienced reader and graded 1 (clear-high level of confidence of interpretation and complete spectral Doppler envelope), 2 (suboptimal with medium-low level of confidence of interpretation and incomplete envelope), 3 (poor-absent and no measurable spectral Doppler signal). Maximal TR velocity (TRV) was defined as peak velocity that could be clearly identified. An inexperienced sonographer read 30 randomly selected studies. Results176 TTE were performed in 173 patients (mean age 57±14.8years). Wilcoxon signed rank test demonstrated significant improvement (p<0.0001) in TR spectral Doppler signal quality with CE TTE. Mean score CE TTE vs. TTE=2.32±0.85 vs. 2.56±0.75 respectively (p<0.0001). Mean maximal TRV CE TTE vs. UE TTE=2.61±0.44m/s vs. 2.54±0.49m/s respectively (p<0.0001). The inexperienced reader had a greater improvement in scoring CE TTE signals vs. UE TTE (p<0.0001). ConclusionIn the era of contemporary scanners, CE improved the ability to detect and measure TRV, except in those with clear unenhanced TR spectral Doppler signals or greater than mild tricuspid regurgitation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call