Abstract
We have embarked upon a systematic program to optimize all aspects of small‐angle‐x‐ray‐scattering x‐ray photon‐correlation spectroscopy experiments performed at beamline 8‐ID‐I at the Advanced Photon Source (APS). We describe two recent efforts, namely better synchronized CCD shutter control and a re‐engineered experiment station layout that has resulted in a 70% increase in optical contrast and a greater than 300% increase in experiment stability. Together these two improvements extend the dynamic range of time scales that can be probed at APS beamline 8‐ID‐I by one decade.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.