Abstract

We evaluated intermittent and continuous thermodilution cardiac output data in 12 patients undergoing orthotopic liver transplantation. Measurements were performed at 16 predefined time points between induction of anaesthesia and 3 h after reperfusion of the liver graft. Cardiac output measurements yielded 192 data pairs (intermittent cardiac output range: 1.8-18.9 l.min-1, continuous cardiac output range: 3.3-20.0 l.min-1). During most of the procedure the correlation between intermittent and continuous cardiac output measurements was significant (r = 0.87, p < 0.0001), accompanied with a bias of -0.240 l.min-1 and a degree of precision of 1.789 l.min-1 (< 10.0 l.min-1:1.137 l.min-1, > or = 10.0 l.min-1:2.220 l.min-1). However, in the early phases after caval clamping and after reperfusion, accuracy was not acceptable. Only during these phases did the difference between the mean values of pulmonary artery blood temperature and rectal temperature increase (after caval clamping) or decrease (after reperfusion). In conclusion, despite acceptable levels of accuracy and precision between intermittent and continuous cardiac output measurement under stable conditions, both methods showed markedly decreased accuracy and precision in the early phases after caval clamping and after reperfusion, possibly owing to increased thermal noise.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.