Abstract

Body temperature must be monitored in patients receiving Hospital-at-Home (HaH) care for COVID-19 and other infectious diseases. Continuous temperature telemonitoring (CTT) detects fever and patient deterioration early, facilitating decision-making. We performed a validation clinical study assessing the safety, comfort, and impact on healthcare practice of Viture®, a CTT system, compared with a standard digital axillary thermometer in 208 patients with COVID-19 and other infectious diseases treated in HaH at the Navarra University Hospital (HUN). Overall, 3258 pairs of measurements showed a clinical bias of -0.02 °C with limits of agreement of -0.96/+0.92 °C, a 95% acceptance rate, and a mean absolute deviation of 0.36 (SD 0.30) °C. Viture® detected 3 times more febrile episodes and revealed fever in 50% more patients compared with spot measurements. Febrile episodes were detected 7.23 h (mean) earlier and modified the diagnostic and/or therapeutic approach in 43.2% of patients. Viture® was validated for use in a clinical setting and was more effective in detecting febrile episodes than conventional methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.