Abstract

The subcutaneous pH values of 42 fetuses has been measured continuously during labor with the use of the electrode designed by Stamm and associates.2 We have obtained 60 per cent good tracings. The results have confirmed the reliability of the subcutaneous pH measurement, if the electrode is correctly inserted, when compared with both capillary and umbilical artery blood pH measurements (r = 0.91, p < 0.001 and 0.83, p <0.001, respectively). Technical improvements appear to be necessary prior to widespread clinical use of continuous pH measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.