Abstract

A measurement system capable of continuous on-line matrix removal, pre-concentration and analysis of 226Ra using pre-packed columns coupled to a flow injection system and an ICP-MS was developed. Full instrumental control of both the ICP-MS and the flow injection system provided automatic integration of the transient signals. The flow injection system was programmed to control column conditioning, sample loading, column rinsing, analyte elution and column cleaning operations employing appropriate solutions. The application of this system to the 226Ra analysis of an industrial liquid effluent was demonstrated. Using this particular instrument together with pre-concentration and matrix removal procedures, a limit of detection of 5.4 fg L−1 (2 mBq L−1) and a method detection limit of 16.2 fg L−1 (6 mBq L−1) were achieved for the measurement of 226Ra using a 25 mL sample volume. Total time for sample handling and analysis is approximately 10 minutes. The concentration of 226Ra in a discharged effluent sample was 0.73 pg L−1 (27 mBq L−1), which is in good agreement with the value of 0.81 pg L−1 (30 mBq L−1) measured using conventional alpha counting techniques.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.