Abstract

BackgroundPatients requiring femoral venoarterial (VA) extracorporeal life support (ECLS) are at risk of distal lower limb hypoperfusion and ischemia of the cannulated leg. In the present study, we evaluated the effect of using continuous noninvasive lower limb oximetry with near-infrared reflectance spectroscopy (NIRS) to detect tissue hypoxia and guide distal perfusion catheter (DPC) placement on the rates of leg ischemia requiring surgical intervention. MethodsWe performed a retrospective analysis of patients who had undergone femoral VA-ECLS at our institution from 2010 to 2014 (pre-NIRS era) and 2017 to 2021 (NIRS era). Patients who had undergone cannulation during the 2015 to 2016 transition era were excluded. The baseline characteristics, short-term outcomes, and ischemic complications requiring surgical intervention (eg, fasciotomy, thrombectomy, amputation, exploration) were compared across the two cohorts. ResultsOf the 490 patients included in the present study, 141 (28.8%) and 349 (71.2%) had undergone cannulation before and after the routine use of NIRS to direct DPC placement, respectively. The patients in the NIRS cohort had had a greater incidence of hyperlipidemia (53.7% vs 41.1%; P = .015) and hypertension (71.4% vs 60%; P = .020) at baseline, although they were less likely to have been supported with an intra-aortic balloon pump before ECLS cannulation (26.9% vs 37.6%; P = .026). These patients were also more likely to have experienced cardiac arrest (22.9% vs 7.8%; P ≤ .001) and a pulmonary cause (5.2% vs 0.7%; P = .04) as an indication for ECLS, with ECLS initiated less often for acute myocardial infarction (15.8% vs 34%; P ≤ .001). The patients in the NIRS cohort had had a smaller arterial cannula size (P ≤ .001) and a longer duration of ECLS support (5 vs 3.25 days; P ≤ .001) but significantly lower rates of surgical intervention for limb ischemia (2.6% vs 8.5%; P = .007) despite comparable rates of DPC placement (49.1% vs 44.7%; P = .427), with only two patients (1.1%) not identified by NIRS ultimately requiring surgical intervention. ConclusionsThe use of a smaller arterial cannula (≤15F) and continuous NIRS monitoring to guide selective insertion of DPCs could be a valid and effective strategy associated with a reduced incidence of ischemic events requiring surgical intervention.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.