Abstract
Fibreoptic jugular bulb oximetry has been validated for use in the care of severely head-injured patients. We compared bench and fibreoptic methods of measuring jugular bulb oxyhaemoglobin saturation (SjO2) in 33 patients undergoing cardiac surgery both during cardiopulmonary bypass (CPB) and in the early postoperative period. After insertion of a fibreoptic reflectance oximetry catheter into the jugular bulb, it was calibrated against a bench oximeter. Comparisons were made while on CPB (n = 60) and in the postoperative period for up to 18 h (n = 215). There was negligible bias throughout. There were wide limits of agreements (mean difference +/- 2SD) between the two methods during operation (-20.29% to 18.05%), whereas after operation the limits of agreement were far narrower (-6.39% and 7.45%). Measurement of SjO2 by the fibreoptic method compared poorly with bench oximetry during CPB but there was good agreement between the two methods in the early postoperative period.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.