Abstract

Accurate assessment and monitoring of the cardiocirculatory function are essential during major pediatric and pediatric cardiac surgery. Monitoring of the central venous oxygen saturation (ScvO(2)) may be a better indicator of tissue oxygenation and derangement of cellular oxygen utilization than the more commonly used vital parameters. Therefore, we compared oxygen saturation measurements with thin fiberoptic oximetry catheters and standard blood gas oximetry in an in vitro setting. Two different size continuous fiberoptic oximetry catheters (2-4-F) were inserted in an extracorporeal circuit filled with human red blood cells in normal saline (haematocrit 30%, flow 600 ml.min(-1)). The results of fiberoptic oximetry were then compared with standard blood gas oximetry for a wide range of different oxygen saturations using linear regression. The oxygen saturations found ranged from 9% to 100%. The results of the two different fiberoptic oximetry catheters correlated significantly (r = 0.99, P < 0.0001) with standard blood gas oximetry. The results of fiberoptic oximetry are nearly identical with standard blood gas oximetry for a wide range of different oxygen saturations. Thin oximetry catheters can be inserted percutaneously even in neonates and small infants. The continuous monitoring of ScvO(2) may be beneficial, especially in patients who are in danger of developing low cardiac output or sudden cardiovascular collapse.