Abstract

Reliable continuous glucose monitoring (CGM) enables a variety of advanced technology for the treatment of type 1 diabetes. In addition to artificial pancreas algorithms that use CGM to automate continuous subcutaneous insulin infusion (CSII), CGM can also inform fault detection algorithms that alert patients to problems in CGM or CSII. Losses in infusion set actuation (LISAs) can adversely affect clinical outcomes, resulting in hyperglycemia due to impaired insulin delivery. Prolonged hyperglycemia may lead to diabetic ketoacidosis—a serious metabolic complication in type 1 diabetes. Therefore, an algorithm for the detection of LISAs based on CGM and CSII signals was developed to improve patient safety. The LISA detection algorithm is trained retrospectively on data from 62 infusion set insertions from 20 patients. The algorithm collects glucose and insulin data, and computes relevant fault metrics over two different sliding windows; an alarm sounds when these fault metrics are exceeded. With the chosen algorithm parameters, the LISA detection strategy achieved a sensitivity of 71.8% and issued 0.28 false positives per day on the training data. Validation on two independent data sets confirmed that similar performance is seen on data that was not used for training. The developed algorithm is able to effectively alert patients to possible infusion set failures in open-loop scenarios, with limited evidence of its extension to closed-loop scenarios.

Highlights

  • Improvement in care for patients living with type 1 diabetes is currently tied to emerging external medical device technology, including continuous glucose monitors and insulin infusion pumps, seeking movement towards closed-loop artificial pancreas systems [1,2]

  • This section defines the criteria for determining an infusion set failure, and follows with a discussion of the data and training procedure used in algorithm development

  • In the derivations that follow, k represents time in minutes, LW represents the length of the long window, SW represents the length of the short window, xi| j represents the average of quantity x in a window from time j to time i, and ∆xk represents the quantity xk − xk−1

Read more

Summary

Introduction

Improvement in care for patients living with type 1 diabetes is currently tied to emerging external medical device technology, including continuous glucose monitors and insulin infusion pumps, seeking movement towards closed-loop artificial pancreas systems [1,2]. Artificial pancreas systems still require patient involvement when changing sensors, choosing infusion set sites, and diagnosing potential problems. One major potential problem with any artificial pancreas system is the loss of the infusion set’s ability to actuate the system. Faulty actuation occurs for a number of reasons, including dislodged catheters, insulin leakage or lipohypertrophy at the infusion set site, and mechanical failure. Alarming a patient when a faulty infusion set is detected has the potential to mitigate or prevent the resulting hyperglycemic glucose excursions

Methods
Results
Conclusion