Abstract

This study aimed to evaluate the effects of posture (sitting [lying down]/standing) on hemodynamic and pump-related parameters in calves implanted with our institution's continuous-flow total artificial heart (CFTAH). These parameters were analyzed with posture information in four calves that had achieved the intended 14-, 30-, or 90-day durations of implantation. In each animal, postoperative hourly data gathered throughout the study were used to compare average values with the animal sitting vs. standing. Pump flow became significantly higher in the standing than sitting position at the same pump speed (standing 7.9 ± 0.8, sitting 7.4 ± 1.0 L/min, p = 0.028). Systemic vascular resistance (SVR) and aortic pressure (AoP) were significantly lower in the standing than sitting position (SVR standing 779 ± 145, sitting 929 ± 206dynes/cm5, p = 0.027; AoP standing 93 ± 7, sitting 103 ± 7mm Hg, p < 0.001). No substantial change occurred in pulmonary vascular resistance (PVR) or pulmonary arterial pressure (PAP) with posture (PVR standing 161 ± 39, sitting 164 ± 48dynes/cm5, p = 0.639; PAP standing 32 ± 3, sitting 33 ± 4mm Hg, p = 0.340). Posture affected some hemodynamic and pump-related parameters in calves with CFTAH, with implications for patients with implanted pumps.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.