Abstract
A microfluidic based device has been developed for the continuous separation of polymer microspheres, taking advantage of the flow characteristics of systems. The chip consists of an asymmetric cavity with variable channel width which enables continuous amplification of the particle separation for different size particles within the laminar flow profile. The process has been examined by varying the sample inlet position, the sample to media flow rate ratio, and the total flow rate. This technique can be applied for manipulating both microscale biological and colloidal particles within microfluidic systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.