Abstract

Acute renal failure can be treated with different dialysis modalities, depending on patient characteristics and hospital resources. Peritoneal dialysis (PD) can be first choice in situations like hypotension, disturbed coagulation, or difficult venous access. The main disadvantage of PD is the relatively limited efficacy. The aim of this study was to investigate whether continuous flow peritoneal dialysis (CFPD) is a more effective treatment than conventional PD in acute renal failure. A pilot study was performed at The Red Cross University Hospital in Cape Town in six patients. Patients were treated with both CFPD and conventional PD for 8 to 16 hours. CFPD was performed with two bedside-placed catheters. After initial filling, dialysate flow rate (100 ml/1.73 m2 per minute) was maintained with an adapted continuous venovenous hemofiltration machine. Ultrafiltration flow rate was set at 2.5 ml/1.73 m2 per minute. Mean ultrafiltration was 0.20 ml/1.73 m2 per minute with conventional PD versus 1.8 ml/1.73 m2 per minute with CFPD. Mean clearances of urea and creatinine were 5.0 and 7.6 ml/1.73 m2 per minute with conventional PD versus 15.0 and 28.8 ml/1.73 m2 per minute with CFPD, respectively. No complications occurred. In this first report of CFPD in six pediatric patients with acute renal failure, CFPD was on average three to five times more effective for urea and creatinine clearance and ultrafiltration than conventional PD, without any complications observed. CFPD has the ability to improve therapy for acute renal failure.