Abstract

It is shown that the iron(III) chalcogenide glass membrane ion-selective electrode (ISE) can be calibrated in continuous flow analysis (CFA) using acidified iron(III) nitrate standards, yielding a 60±3 mV per decade change in activity of Fe 3+ response in the range 10 −7–10 −2 M total iron(III). Extended ageing of the iron(III) ISE in 2 M zinc(II) sulphate did not alter the potentiometric response characteristics of the electrode. Furthermore, electrochemical impedance spectroscopy and X-ray photoelectron spectroscopy in the presence and absence of zinc(II) sulphate failed to detect a zinc(II) interference on the iron(III) ISE. CFA/ISE determined activities of Fe 3+ in synthetic zinc electrolyte containing 2×10 −3–2×10 −1 M total iron(III) yielded results falling within ±0.2 log aFe 3+ unit of the corresponding iron speciation data calculated using the minteqa2 program.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.