Abstract

A system for the detection of radioactive effluents from a gas chromatograph is described. The effluents are continuously dissolved in a stream of scintillator solution. After most of the carrier gas has been split off, the solution is led at a constant rate through a helical flow cell inserted in a liquid scintillaiton counter. The main advantages of the system are its expediency for all radioactive compounds, the absence of tailing or memory effects, good resolution and high sensitivity (less than 0.5 nCi). Peak fractions can be collected after they have passed through the flow cell for accurate activity measurements. The method is also suitable for distinguishing between compounds labelled with two different radioactive isotopes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.