Abstract

We have analysed the clinical agreement between two methods of continuous cardiac output measurement pulse contour analysis (PCCO) and a continuous thermodilution technique (CCO), were both compared with the intermittent bolus thermodilution technique (BCO). Measurements were performed in 26 cardiac surgical patients (groups 1 and 2, 13 patients each, with an ejection fraction > 45% and < 45%, respectively) at 12 selected times. During operation, mean differences (bias) between PCCO-BCO and CCO-BCO did not differ in either group. However, phenylephrine-induced increases in systemic vascular resistance (SVR) by approximately 60% resulted in significant differences. Significantly higher absolute bias values of PCCO-BCO compared with CCO-BCO were also found early after operation in the ICU. Thus PCCO and CCO provided comparable measurements during coronary bypass surgery. After marked changes in SVR, further calibration of the PCCO device is necessary.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.