Abstract
An ammonia monitor has been made by using a rectangular Stark microwave cavity tightly coupled through an iris to a Gunn oscillator. The Stark electrode installed inside the cavity is provided with a 100 kHz sinusoidal modulation voltage and a dc sweep voltage. The oscillation of the Gunn diode is tuned with the coupled Stark cavity to a frequency near a J,K=3,3 transition (23870.1 MHz) of ammonia. By fixing the Stark dc bias to a voltage which gives a peak intensity of the derivative output of the electric-resonance signal, ammonia is continuously monitored. An extremely good long-term stability is obtained. The noise level corresponds to 0.08 ppm of ammonia.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.