Abstract
We have developed a system for contactless measurement of nonlinear conductivity in the radio-frequency band, and over a wide temperature range. A non-resonant circuit is used to electrically excite the sample, and the induced signal is detected by a resonant circuit whose natural frequency matches higher harmonics of the excitation. A simple modification of the probe allows non-resonant detection suitable for stronger signals. Two measurement procedures are proposed that allow significant excitation power variation, up to 150 W. The apparatus has been validated through the measurement of the nonlinear response at the superconducting transition of a high-Tc superconductor, and the nematic transition of an iron pnictide.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.