Abstract

Background Unicompartmental knee replacement has gained popularity in recent times, showing improved success rates. The main reasons for the failure of unicompartmental knee replacement are the wear of the polyethylene bearing, aseptic loosening and wear in the opposite compartment. The contact stresses involved are significant contributing factors to these causes of failure. Methods In this study, a patient-specific unicompartmental knee replacement is proposed using a methodology based on neural network modeling of a database of healthy knee geometries. This custom implant was then compared to two conventional implant designs in terms of contact stress in a validated finite element model. Findings The custom implant experienced lower contact stresses at the tibio-femoral joint compared to a fixed-bearing design and also displayed more uniform stress distribution at the bone–implant interface than any of the other implant designs. Interpretation Custom unicompartmental knee replacements therefore have the potential of providing good contact stress distribution, preserve bone stock and could be more anatomically accurate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.