The development of computational and numerical methods provides the option to study the contact mechanics and wear of hip resurfacing devices. The importance of these techniques is justified by the extensive amount of testing and experimental work required to verify and improve current orthopaedic implant devices. As the demands for device longevity is increasing, it is as important as ever to study techniques for providing much needed orthopaedic hip implant solutions. Through the use of advanced computer aided design and the finite element method, contact analysis of hip resurfacing devices was carried out by developing both three-dimensional and two-dimensional axisymmetric models whilst considering the effects of loading conditions and material properties on the contact stresses. Following on from this, the three-dimensional model was used in combination with a unique programme to develop wear simulations and obtain cumulative wear for both the acetabular cup and femoral head simultaneously.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.