Abstract
Complications, although rare, do occur with cochlear implantation. We present a 2-year old with persistent erythema and pruritus over the implantation site with serous discharge from the implanted ear. Patch testing revealed contact dermatitis to silicone LSR-30 encasing the Nuclear Contour-24 device. Hypersensitivity was also noted to silicone LSR-30 in the Med-el device but not to silicone LSR-70 in the Advance Bionics device. Device explantation resulted in complete resolution of symptoms. The patient was then successfully reimplanted with the Advance Bionics device. Erythema and irritation at the implantation site, without leukocytosis, should prompt evaluation for silicone contact dermatitis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.