Abstract

We have constructed an Ultra Low Temperature cryostat for Magnetic Resonance Imaging (ULT-MRI) applicable for various condensed phases of 3He at ultra low temperature, such as Nuclear Ordered Solid 3He or superfluid 3He. In contrast to conventional MRI for general use, where a pulsed-field gradient method is commonly used, we used a steady-field gradient method to avoid eddy current heating to metal parts at ultra low temperature. To avoid nonlinear response from the nuclear ordered spin system, which contradicts a principle of conventional MRI, we employed Free Induction Decay (FID) after a small tipping angle RF pulse to obtain the spectrum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.