Abstract
An isoperibolic micro-combustion calorimeter was designed, built and set up in our laboratory, taking as base a 1107 Parr combustion bomb of 22 cm 3 of volume. Taken into account the geometrical form of the bomb, it was designed and constructed a vessel and a submarine chamber in brass. All of the pieces of the calorimeter were chromium-plated to reduce heat loss by radiation. The calorimeter was calibrated by using pellets of standard benzoic acid (mass approximate of 40 mg) leading to the energy equivalent of ε(calor) = (1283.8 ± 0.6) J · K −1. In order to test the calorimeter, combustion experiments of salicylic acid were performed leading to a value of combustion energy of Δ c u ∘ = −(21,888.8 ± 10.9) J · g −1, which agrees with the reported literature values. The combustion of piperonylic acid was carried out as a further test leading to a value of combustion energy of Δ c u ∘ = −(20,215.9 ± 10.4) J · g −1 in accordance with the reported literature value. The uncertainty of the calibration and the combustion of salicylic acid and piperonylic acid was 0.05%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.