Abstract

A picrate ion-selective electrode without inner reference solution and based on bis(triphenylphosphoranylidene)ammonium picrate dissolved in 2-nitrophenyloctyl ether was constructed and its characteristics assessed. With respect to reducing sugar determination in Port wine, a tubular electrode based on the same sensor system was developed and incorporated into a flow-injection manifold devised for this determination. The methodology involves a reaction between the reducing sugar and picric acid, the decrease in the picrate concentration being monitored with the tubular electrode. A linear dependence between peak width, at a fixed potential and the logarithm of sugar concentration in the range 25–200 g l −1 was obtained. The proposed system allows wine samples to be injected with no previous treatment and allows a sample throughput of 50 determinations per hour, a relative standard deviation less than 4%, and the results are comparable to the reference procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.