Abstract
At the University of Twente, a heart scanner has been designed and constructed that uses superconducting devices (superconducting quantum interference devices (SQUIDs)) to measure the magnetic field of the heart. A key feature is the elimination of liquid cryogens by incorporating cryocoolers. In the design, two coolers are operated in counter-phase to reduce the mechanical interference. In addition to the application of ferromagnetic shields around the compressors, the magnetic cooler interference is reduced by placing the SQUID magnetometers coplanar with respect to the coolers. In this way, the cooler noise was reduced to a level below the intrinsic sensor noise: 0.16 pT/√Hz. A temperature of 60 K was realised with a cool-down time of about 2 h. The corresponding heat load to the coolers is roughly 0.9 W. Magnetocardiograms were recorded inside a magnetically shielded room.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.